FDA recall Z-3038-2017

Boston Scientific Corporation · Class II · device

Product

EMBLEM MRI S-ICD Model 219, Subcutaneous-Implantable Cardioverter Defibrillator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing

Reason for recall

The device can deliver an atypical amount of energy due to memory corruption inside the device.

Distribution

Worldwide - US Nationwide distribution, including Puerto Rico, U.S. Virgin Island, and Guam, There was also worldwide foreign distribution, including Canada.

Key facts

Status
Terminated
Initiation date
2017-06-29
Report date
2017-08-23
Termination date
2019-02-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Paul, MN, United States