FDA recall Z-3040-2020

Ameditech Inc · Class II · device

Product

iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300) 5 Adulteration Tests: Cr, Ni, pH, Bl, S.G.", X11-I-DX-1147-022 /"iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300)"

Reason for recall

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Distribution

US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Key facts

Status
Terminated
Initiation date
2020-08-18
Report date
2020-10-07
Termination date
2024-09-30
Voluntary/Mandated
Voluntary: Firm initiated