FDA recall Z-3051-2020

Ameditech Inc · Class II · device

Product

Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Sheet-Card /187876, 100195/ EDI PPX Uncut Sheet-Card/ 187877, 160130/ EDI COC Uncut Sheet-Cup/ 188168, 160180/ EDI PCP Uncut Sheet-Cup/ 189027, 160140/ EDI THC Uncut Sheet Cup /189028, 160170/ EDI BZO Uncut Sheet-Cup/ 189081, 160120/ EDI MET1000 Uncut Sheet-Cup/ 189082, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189123, 160130/ EDI COC Uncut Sheet-Cup/ 189124, 160160/ EDI OPI2000 Uncut Sheet-Cup/ 189166, 160163/ EDI OPI (300) Uncut Sheet-Cup/ 189191, 160110/ EDI AMP Uncut Sheet-Cup/ 189371, 160140/ EDI THC Uncut Sheet Cup/ 189372, 160123/ EDI MET(300) Uncut Sheet-Cup/ 189425, 160170/ EDI BZO Uncut Sheet-Cup/ 189426, 100163/ EDI OPI (300) Uncut Sheet-Card/ 189456, 160170/ EDI BZO Uncut Sheet-Cup/ 189565, 160180/ EDI PCP Uncut Sheet-Cup/ 189601, 160125/ EDI MET(500) Uncut Sheet-Cup/ 189748, 160130/ EDI COC Uncut Sheet-Cup/ 189786, 160160/ EDI OPI2000 Uncut Sheet-Cup /

Reason for recall

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Distribution

US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Key facts

Status
Terminated
Initiation date
2020-08-18
Report date
2020-10-07
Termination date
2024-09-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3051-2020