FDA recall Z-3053-2020

ESAOTE S.P.A · Class II · device

Product

AC/DC POWER SUPPLY for 7400, Part Number 151004301, XP Power Model Number AHM180PS19-XE1057, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005)

Reason for recall

The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.

Distribution

US Nationwide

Key facts

Status
Terminated
Initiation date
2020-09-08
Report date
2020-10-07
Termination date
2023-04-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Genova, N/A, Italy

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3053-2020