FDA recall Z-3065-2024

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;

Reason for recall

During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.

Distribution

Domestic: Nationwide Distribution. International: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Qatar, Slovenia, Spain, Switzerland.

Key facts

Status
Ongoing
Initiation date
2024-07-30
Report date
2024-09-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Newark, DE, United States