# FDA recall Z-3069-2017

> **Medtronic Neuromodulation** · Class II · device recall initiated 2017-03-17.

## Product

DBS Lead Depth Stop contained in Medtronic Deep Brain Stimulation (DBS) lead kits, models 3387/3387S, 3389/3389S, and 3391/3391S,  The lead kit contains one DBS lead in addition to multiple accessories used in DBS lead implant procedures. Among these accessories is the DBS lead holder, also referred to as the lead depth stop.   The lead holder affixes to the DBS lead during the lead implant procedure, marking the distance to the lead tip and providing control of DBS lead depth during implant by interfacing with the stereotactic system

## Reason for recall

Medtronic received reports that the DBS depth stop did not adequately secure to the lead, which can result in DBS lead placement beyond the intended target.

## Distribution

Worldwide Distribution - US including AL, AR, AZ, CA,CO, CT, DC, FL,  GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME,  MI, MN,M MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.    Foreign Distribution: India, ANZ, Taiwan, Hong Kong, Canada, China, Japan, Korea, Malaysia, Philippines, Singapore, Thailand, Indonesia, Vietnam, Albania, Algeria, Austria, Belarus, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, India, Iran, Iraq, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, United Kingdom, Argentina, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Mexico, Panama, Peru, Puerto R

## Key facts

- **Recall number:** Z-3069-2017
- **Recalling firm:** Medtronic Neuromodulation
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-03-17
- **Report date:** 2017-08-30
- **Termination date:** 2021-04-22

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Minneapolis, MN, United States