FDA recall Z-3103-2024

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

ADVIA Centaur CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491379 SMN: 11206239

Reason for recall

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

Distribution

Worldwide - US Nationwide distribution

Key facts

Status
Ongoing
Initiation date
2024-07-26
Report date
2024-09-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
East Walpole, MA, United States