FDA recall Z-3114-2024

MEDLINE INDUSTRIES, LP - Northfield · Class II · device

Product

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number DYKE1092B; b) VIDEO BRONCHOSCOPY TRAY, Pack Number DYKE1666; c) SUCTION CANISTER KIT, Pack Number DYKM2196; d) SUCTION CANISTER KIT, Pack Number DYKM2196H; e) BRONCH LAB PACK-LF, Pack Number DYNJ24810B; f) THORACIC, Pack Number DYNJ907991F; g) RECTAL, Pack Number DYNJ908108A; h) ROBOTIC, Pack Number DYNJ908109C; i) CERVICAL, Pack Number DYNJ908748D; j) LAMI, Pack Number DYNJ908750D; k) MERCY TIFFIN BASIC-LF, Pack Number DYNJ909731; l) MERCY TIFFIN LITHOTOMY, Pack Number DYNJ909735; m) BASIC GENERAL SET UP, Pack Number DYNJ909945; n) BASIC GENERAL SET UP, Pack Number DYNJ909945A; o) AMB OCULOPLASTICS MUSC, Pack Number DYNJ910289; p) AMB OCULOPLASTICS MUSC, Pack Number DYNJ910289H; q) MTHZ TOTAL HIP, Pack Number DYNJC9311M; r) BRONCH KIT, Pack Number P906877

Reason for recall

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Key facts

Status
Ongoing
Initiation date
2024-06-27
Report date
2024-09-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3114-2024