FDA recall Z-3136-2024

Datascope Corp. · Class I · device

Product

CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

Reason for recall

Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.

Distribution

International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, IRAQ, IRAN, ITALY, JORDAN, JAPAN, SOUTH KOREA, KUWAIT, LEBANON, LITHUANIA, LIBYA, MOROCCO, MYANMAR, MEXICO, MALAYSIA, NETHERLANDS, NORWAY, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PALESTINE, PORTUGAL, SERBIA, RUSSIAN FEDERATION, SAUDI ARABIA, SWEDEN, SINGAPORE, SLOVAKIA, THAILAND, TURKEY, TAIWAN, UNITED KINGDOM, VIETNAM, YEMEN.

Key facts

Status
Ongoing
Initiation date
2024-08-08
Report date
2024-10-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mahwah, NJ, United States

Primary source