FDA recall Z-3141-2017

Medtronic Navigation, Inc. · Class II · device

Product

Spine Referencing Instrumentation, Double Spinous Process Clamp, Tall When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.

Reason for recall

Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.

Distribution

Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, AD, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, & WY. Foreign: Albania, Australia, Bangladesh, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Kenya, Latvia, Luxembourg, Malaysia, Martinique, Netherlands, New Zealand, Philippines, Poland, Portugal, Romania, Russia, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, & Vietnam.

Key facts

Status
Terminated
Initiation date
2017-06-29
Report date
2017-09-27
Termination date
2023-07-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Louisville, CO, United States

Primary source