Herbert Cannulated Bone Screw System, Cannulated Main Drill 3.1 mm Diameter, Item Number/EDI 00115501500, Nonsterile
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3145-2017