FDA recall Z-3160-2024

GE Healthcare · Class II · device

Product

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 2110344-009; c) 2110344-010; d) 2110344-011; e) 2110344-013; f) 2110344-019; g) 2110344-025; h) 2110344-029; i) 2110344-030; j) 2110344-032 ; Radiological image processing system

Reason for recall

Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2024-07-23
Report date
2024-09-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Chicago, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3160-2024