FDA recall Z-3166-2024

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Class II · device

Product

System Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 OR Table 722038 Allura Xper FD20/20 722035 Allura Xper FD20 OR Table 722029 Allura Xper FD20/10 722028 Allura Xper FD20 722027 Allura Xper FD10/10 722026 Allura Xper FD10 722025 Allura Xper FD20 Biplane OR Table 722023 Allura Xper FD20 OR Table 722020 Allura Xper FD20 Biplane OR Table 722015 Allura Xper FD20 OR Table 722013 Allura Xper FD20 Biplane 722012 Allura Xper FD20 722011 Allura Xper FD10/10 722010 Allura Xper FD10 722008 Allura Xper FD20 Biplane 722006 Allura Xper FD20 722005 Allura Xper FD10/10 722003 Allura Xper FD10 722002 Allura Xper FD10F 722001 Allura Xper FD10C

Reason for recall

Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

Distribution

Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R¿union Romania Russian Federation Saud

Key facts

Status
Ongoing
Initiation date
2024-08-22
Report date
2024-09-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Best, Netherlands

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3166-2024