FDA recall Z-3175-2024

B Braun Medical Inc · Class II · device

Product

STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095DA

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2024-08-15
Report date
2024-09-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bethlehem, PA, United States