FDA recall Z-3187-2017

Zimmer Biomet, Inc. · Class II · device

Product

ECT Internal Fracture Fixation Quick-Connect Drill 3.5 mm Diameter 11 cm Length, Item Number/EDI 00241003500, Nonsterile. orthopedic surgical instrument.

Reason for recall

Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.

Distribution

Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.

Key facts

Status
Terminated
Initiation date
2017-05-22
Report date
2017-10-04
Termination date
2019-03-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3187-2017