FDA recall Z-3188-2018

EOS Imaging · Class II · device

Product

sterEOS workstation

Reason for recall

3D projections may be incorrect when the pair of images used is a secondary capture generated from a series containing more than two images, where one of the images has been manually flipped. Projection errors may lead to errors in calculation of 3D clinical parameters when adjusting the position, size and shape of the 3D objects used. Incorrect clinical parameters can be a contributing factor in incorrect clinical decisions, or sub-optimal surgical interventions.

Distribution

US nationwide

Key facts

Status
Terminated
Initiation date
2018-08-01
Report date
2018-09-26
Termination date
2022-08-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Paris, N/A, France