FDA recall Z-3195-2018

Fresenius Medical Care Renal Therapies Group, LLC · Class II · device

Product

2008T HEMODIALYSIS SYSTEM W/BIBAG, Part Number 190766, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.

Reason for recall

In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.

Distribution

US Nationwide in the states of : CA, CT, MI, NC, and NY.

Key facts

Status
Terminated
Initiation date
2018-07-13
Report date
2018-09-26
Termination date
2019-09-17
Voluntary/Mandated
Voluntary: Firm initiated