FDA recall Z-3248-2024

Hologic Inc · Class II · device

Product

Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.1 SYSTEMS 2. ASY-16825 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.1 SYSTEMS 3. ASY-17219 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.10.0 SYSTEMS 4. ASY-17220 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.1.0 SYSTEMS 5. ASY-17221 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.0 SYSTEMS 6. ASY-17222 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.0 SYSTEMS 7. 3DM-LIC-SW-UP-V2.3 DIMENSIONS SOFTWARE 2.3 (3DIMENSIONS) 8. SDM-LIC-SW-UP-V1.12 DIMENSION SOFTWARE 1.12 (SELENIA DIMENSIONS)

Reason for recall

Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis

Distribution

Worldwide distribution - US Nationwide and the countries of Belgium, Canada, Israel, Korea, Republic of, Netherlands.

Key facts

Status
Ongoing
Initiation date
2024-08-19
Report date
2024-10-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Newark, DE, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3248-2024