FDA recall Z-3312-2024

Philips North America Llc · Class II · device

Product

DIVA 24 Inch Widescreen LCD Touch Display - Used with the Philips PIC iX system central monitoring system. Model: RGD2461AMI Philips Part # 866126

Reason for recall

If the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a patient condition which prompted the alarm to generate. Although unlikely, this could potentially result in a patient harm.

Distribution

Worldwide - US Nationwide distribution in the states of CA, IL, ME, NC, NJ, NM, NY, OH, VA, WA and the countries of Australia, Austria, Canada, France, Germany, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2024-02-15
Report date
2024-10-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cambridge, MA, United States