FDA recall Z-0877-2020

GE Healthcare, LLC · Class II · device

Product

SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

Reason for recall

It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.

Distribution

Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2019-12-23
Report date
2020-02-05
Termination date
2020-12-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source