FDA recall Z-1256-2022

Steris Corporation · Class II · device

Product

Reliance Synergy Washer/Disinfector

Reason for recall

The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

Distribution

Worldwide distribution - US Nationwide.

Key facts

Status
Terminated
Initiation date
2022-04-22
Report date
2022-06-15
Termination date
2024-11-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mentor, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1256-2022