FDA recall Z-1294-2019

Medica Corporation · Class II · device

Product

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.

Reason for recall

Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.

Distribution

Worldwide distribution - US Nationwide and countries of IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE GUATEMALA, KENYA, IRAN, ESTONIA.

Key facts

Status
Terminated
Initiation date
2019-02-06
Report date
2019-05-15
Termination date
2020-10-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bedford, MA, United States

Primary source