FDA recall Z-1577-2025

Maquet Cardiovascular, LLC · Class II · device

Product

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

Reason for recall

Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.

Distribution

Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.

Key facts

Status
Ongoing
Initiation date
2025-03-10
Report date
2025-04-23
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States