FDA recall Z-1729-2025

Remote Diagnostic Technologies Ltd. · Class II · device

Product

Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007

Reason for recall

Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.

Distribution

Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.

Key facts

Status
Ongoing
Initiation date
2025-04-15
Report date
2025-05-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Farnborough, N/A, United Kingdom

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1729-2025