FDA recall Z-1777-2019

Maquet Cardiovascular Us Sales, Llc · Class II · device

Product

Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.

Reason for recall

Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.

Distribution

Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine

Key facts

Status
Terminated
Initiation date
2019-05-20
Report date
2019-06-19
Termination date
2023-12-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States

Primary source