FDA recall Z-1921-2024

Thoratec Corp. · Class I · device

Product

Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device

Reason for recall

System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected

Distribution

Worldwide distribution. US nationwide including Puerto Rico, UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, BULGARIA, SWITZERLAND, CYPRUS, CZECH REPUBLIC, GERMANY, DENMARK, ESTONIA, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, GUADELOUPE, GREECE, IRELAND, ISRAEL, IRAN, ISLAMIC REPUBLIC OF, ITALY, KUWAIT, LEBANON, LITHUANIA, LUXEMBOURG, MACEDONIA, MARTINIQUE, NETHERLANDS, NORWAY, OMAN, POLAND, PALESTINE, PORTUGAL, QATAR, ROMANIA, SERBIA, RUSSIAN FEDERATION, SWEDEN, SLOVENIA, TUNISIA, TURKEY, SOUTH AFRICA, CANADA, AUSTRALIA, HONG KONG, MALAYSIA, SINGAPORE, SOUTH KOREA, TAIWAN, THAILAND, JAPAN, BANGLADESH, INDIA, PAKISTAN, ARGENTINA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, MEXICO, and BRAZIL

Key facts

Status
Ongoing
Initiation date
2024-05-08
Report date
2024-06-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Pleasanton, CA, United States