FDA recall Z-1974-2023

QUIDEL CARDIOVASCULAR INC · Class I · device

Product

Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.

Reason for recall

Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.

Distribution

US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD. OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL

Key facts

Status
Ongoing
Initiation date
2023-05-25
Report date
2023-07-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States