FDA recall Z-2032-2025

Baxter Healthcare Corporation · Class I · device

Product

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Reason for recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Distribution

Worldwide

Key facts

Status
Ongoing
Initiation date
2025-05-30
Report date
2025-07-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2032-2025