FDA recall Z-2545-2024

Beckman Coulter, Inc. · Class II · device

Product

Discrete photometric chemistry analyzer for clinical use.

Reason for recall

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

Distribution

Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.

Key facts

Status
Ongoing
Initiation date
2024-06-14
Report date
2024-08-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Chaska, MN, United States