FDA recall Z-2747-2020

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US

Reason for recall

If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.

Distribution

Nationwide

Key facts

Status
Terminated
Initiation date
2020-06-30
Report date
2020-08-12
Termination date
2020-11-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Newark, DE, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2747-2020